Burette: Uses and How to Use It

Home › Laboratory Equipment

Laboratory equipment guide

Burette: Uses and How to Use It

A burette is a long graduated glass tube with a stopcock at the bottom, used to deliver accurately measured volumes of solution, mainly in titration. Its scale starts at 0 at the top, so the volume delivered = final reading − initial reading.

Parts of a burette

What is a burette used for?

  • Adding titrant gradually in acid-base, redox or complexometric titrations.
  • Delivering accurate volumes (typically to 0.05 mL for a 50 mL burette).
  • Determining the concentration of an unknown solution.

Parts of a burette

  1. 1Zero at the top – the scale runs down to 25 or 50 mL
  2. 2Graduated tube – lines every 0.1 mL on a 50 mL burette
  3. 3Stopcock – glass or PTFE, controls the flow
  4. 4Tip – where the drops come out
Parts of a burette: 1) Zero at the top, 2) Graduated tube, 3) Stopcock, 4) Tip.
Parts of a burette: 1) Zero at the top, 2) Graduated tube, 3) Stopcock, 4) Tip.

How to use a burette

Burette – Step 1: Clamp it vertically on a stand
Step 1

Clamp it vertically on a stand

Clamp the burette to a retort stand with a burette clamp. Make sure it is perfectly vertical and the stopcock is closed.

Burette – Step 2: Fill with a funnel, remove bubbles
Step 2

Fill with a funnel, remove bubbles

Rinse the burette with a little titrant. Fill it through a small funnel (1) above the zero mark, remove the funnel, then open the stopcock briefly into a waste beaker (2) to clear bubbles from the tip and bring the level to 0.00 mL or another recorded value.

Burette – Step 3: Titrate: left hand on the stopcock, right hand swirling
Step 3

Titrate: left hand on the stopcock, right hand swirling

Work the stopcock with your left hand (1), wrapped around the burette from behind so the tap is not pulled out. Swirl the flask with your right hand (2). Near the end point, add drop by drop until the indicator colour changes and persists.

Burette – Step 4: Read the bottom of the meniscus at eye level
Step 4

Read the bottom of the meniscus at eye level

Read the initial and final volumes at eye level with the bottom of the meniscus, to two decimal places (e.g. 23.45 mL). A dark card behind the burette makes the meniscus easier to see.

Common mistakes

  • Leaving an air bubble in the tip: the recorded volume is larger than what was delivered.
  • Reading from above or below, or forgetting that the scale increases downwards.
  • Leaving the funnel in place during titration: residual drops fall in.

Care and safety

  • Empty and rinse after use, especially after alkalis that can seize glass stopcocks.
  • Store upside down with the stopcock open in a burette rack.

SOP for using a burette

A summary of the standard operating procedure (SOP) for a burette that your laboratory can adapt. Each step is explained in the how-to section above.

1Protective equipment

  • A buttoned, long-sleeved lab coat.
  • Safety glasses; goggles where corrosive splashes are possible.
  • Nitrile gloves suited to the chemicals handled; heat-resistant gloves for hot items.

2Before use

  • Check for cracks, chips and deep scratches. Do not use damaged glassware; set it aside and discard it in the broken-glass bin.
  • Test the stopcock: fill with water, close the tap and wait 5 minutes; the tip must stay dry and the level must not fall.
  • Rinse the burette 2–3 times with a little titrant before filling.
  • Read the SDS of every chemical you will handle.

4After use

  • Drain the burette into waste; never return leftover titrant to the stock bottle.
  • Rinse with distilled water, including through the stopcock.
  • Store upside down in a burette rack with the stopcock open. Grease glass stopcocks lightly if needed; PTFE stopcocks need none.

5Verification & calibration

  • Gravimetric check: weigh the clean, dry item, fill or deliver room-temperature distilled water as in normal use, weigh again and convert the mass of water to volume with the Z-factor (about 1.0029 mL/g at 20 °C). Repeat at least 3 times and compare the mean with the tolerance.
  • For burettes, check several points (e.g. 10, 20, 30, 40 and 50 mL). Class A tolerances (ISO 385): 50 mL ±0.05 mL; 25 mL ±0.03 mL.
  • Do this on receipt and then periodically (e.g. every 6–12 months) or after the item has been overheated.

6Emergencies

  • Breakage: never pick up shards by hand. Use a brush and dustpan or tongs and put them in the broken-glass bin.
  • Chemical spill: warn people nearby, secure the area and clean up with a spill kit according to the SDS.
  • Chemical on skin or in eyes: rinse with running water for at least 15 minutes (eyewash or safety shower), then report to the lab supervisor.

A print-ready version with purpose, scope, illustrated procedure, verification and sign-off boxes. Fill in your own document number and effective date.

Download the SOP (PDF, A4)

Types and sizes

Common sizes are 10, 25 and 50 mL, with microburettes for small-scale titrations. PTFE stopcocks need no grease and resist alkali; amber burettes are used for light-sensitive titrants such as KMnO₄ and AgNO₃. Class A is used for quantitative analysis.

Burette at AMI Scientific

Examples of burette available in our store. Prices exclude VAT.

AMI Scientific supplies laboratory glassware, instruments and chemicals with tax invoices and delivery across Indonesia.

Frequently asked questions

Why is the zero at the top of a burette?

Because a burette measures the volume delivered. The liquid level falls as titrant runs out, so readings increase downwards.

How do you calculate the titrant volume?

Titrant volume = final reading − initial reading. Calculate concentration with the titration formula M₁V₁ = M₂V₂ (adjusted for the reaction ratio).