Cyclic Voltammetry Guide
Cyclic Voltammetry: Principles, Reading the Curve and Measurement Tips
Cyclic voltammetry (CV) is the most widely used electrochemical technique. This guide explains how it works, how to read a voltammogram, the key equations and the mistakes that most often make CV data unreliable.
What is cyclic voltammetry?
In CV, the potentiostat sweeps the working electrode potential linearly from a start value to a vertex and back while measuring current. The plot of current against potential is the voltammogram. When the potential passes the redox potential of a species, it is oxidised or reduced and a current peak appears.
The shape of the voltammogram tells you a lot: peak positions give redox potentials, peak separation indicates how fast electrons transfer, peak height relates to concentration and diffusion coefficient, and a rectangular curve indicates capacitive behaviour as in supercapacitors.
How to read a voltammogram
| Parameter | Meaning | Reference value |
|---|---|---|
| Epa, Epc | Anodic and cathodic peak potentials | E½ ≈ (Epa + Epc)/2 approximates the formal potential |
| ΔEp = Epa − Epc | Electron-transfer rate | ≈ 57–59/n mV at 25 °C for a reversible system |
| ipa / ipc | Stability of the reaction product | ≈ 1 for a reversible system without follow-up reactions |
| ip versus √ν | Linear means diffusion-controlled | Randles–Sevcik equation |
| ip versus ν | Linear means adsorbed or capacitive species | Common for modified electrodes |
| Rectangular curve | Electric double-layer charging | Carbon supercapacitors |
Key equations
ip = 2,69 × 10⁵ · n^(3/2) · A · D^(1/2) · C · ν^(1/2)ip in A, A = electrode area (cm²), D = diffusion coefficient (cm²/s), C = concentration (mol/cm³), ν = scan rate (V/s). Used to determine active electrode area with ferri/ferrocyanide.
i = a · ν^bb ≈ 0.5 means diffusion-controlled (battery-like); b ≈ 1 means capacitive (supercapacitor-like).
ΔEp ≈ 2,22 · RT / (nF) ≈ 57/n mV (25 °C)Larger values indicate slow electron transfer (quasi-reversible) or uncompensated solution resistance.
Steps for a good CV measurement
- Prepare the electrodesPolish the working electrode with alumina, rinse and sonicate briefly. Check the reference electrode has no bubbles.
- Remove oxygenPurge with nitrogen or argon for 10–15 minutes when measuring reduction; oxygen adds extra peaks.
- Check with a standard systemMeasure 5 mM K₃[Fe(CN)₆] in 0.1 M KCl; ΔEp of about 60–80 mV means the setup is healthy.
- Choose the potential windowStart narrow; avoid the solvent's hydrogen and oxygen evolution regions.
- Vary the scan rateMeasure at several scan rates (e.g. 10, 25, 50, 100, 200 mV/s) to reveal the mechanism.
- Report full conditionsState the reference electrode, electrolyte, scan rate, electrode area and which cycle is shown.
Common mistakes
- Not correcting solution resistance (iR drop), which broadens and shifts the peaks.
- Showing the first, unstable cycle.
- A clogged reference electrode or dried-out filling solution.
- Reporting potentials without naming the reference electrode.
- Comparing currents without normalising by area or mass.
- A window so wide that the electrolyte decomposes.
Equipment for CV
Every Corrtest potentiostat runs CV, including models without EIS such as the CS300M. For modified electrodes and sensors use a glassy carbon electrode, Ag/AgCl and an electrode polishing kit. A model guide by application is on the electrochemistry page; a real CV example is in the Fe₃O₄/CQDs case study.
Frequently asked questions
What is the difference between CV and LSV?
LSV sweeps the potential in one direction only; CV sweeps forward and back so oxidation and reduction appear in one curve.
What scan rate should I use?
50–100 mV/s for a first survey. Use a range of scan rates for mechanism studies. Battery materials often use 0.1–1 mV/s to stay near equilibrium.
Why is my ΔEp much larger than 60 mV?
Usually a dirty electrode surface, high solution resistance, or genuinely slow electron transfer. Re-polish the electrode and enable iR compensation.
Can AMI run CV on my samples?
Yes, through our electrochemical analysis service.
Want to run CV on your samples?
Use our electrochemical analysis service, or ask us about the right potentiostat and electrode configuration.
Related topics
- Electrochemistry
- Corrosion & Coating Testing
- Battery & Supercapacitor Research
- Electrocatalysis & Fuel Cells
- Electrochemical Sensors & Biosensors
- Heavy-Metal Analysis (Stripping)
- Electrodeposition & Electroplating
- EIS (Impedance) Guide
- Tafel & Corrosion Rate Guide
- Potentiostat Training
- Potentiostat prices 2026
- Electrochemical accessories