The Ossila UV Ozone Cleaner is a benchtop instrument for removing organic contamination from substrates and samples. Its UV lamps emit at 185 nm and 254 nm to produce ozone, which breaks surface contaminants down into volatile compounds that evaporate, leaving near-atomically clean surfaces without damaging the sample.
The drawer tray offers a 153 mm (6") diameter cleaning area and accepts samples up to 12 mm (0.47") high, so several samples, microscope slides, diced substrates, petri dishes or AFM tips can be treated in one run. Exposure is digitally set from 1 second to 60 minutes, and a 4.3" colour LCD with tactile buttons shows temperature readings for the tray, lamp and electronics so delicate samples are kept from overheating.
Safety measures include a hardware drawer interlock that keeps the lamps off while the tray is open, active lamp cooling and a software thermal cutoff. The unit carries the CE mark and meets BS EN 61010-1:2010. The system weighs 5.25 kg and accepts 110 V – 240 V input, with US, EU, UK and AU variants.
Related products
- Ossila Solar Cell I-V Test System without Solar Simulator T2002F2
- Ossila Four Point Probe Original Soft-tipped probe head T2001A4
- Ossila Contact Angle Goniometer
- Ossila Dip Coater L2006A2-US/EU/UK/AU
- Ossila LED Solar Simulator G2009B2
- Ossila Potentiostat None Electrochemical Cell and Electrodes T2006A1-US/EU/UK/AU
- Removing organic contaminants from substrate surfaces before further experimental steps.
- Cleaning microscope slides and diced substrates.
- Treating petri dishes inside the cleaning tray.
- Cleaning AFM tips without damaging them.
- Preparing wafers up to 153 mm (6") in diameter.
| Brand | Ossila |
|---|---|
| CAS number | — |
| Molecular formula | — |
| Purity | — |
| Category | — |
| Pack sizes | — |
| Physical form | — |
| Storage | — |
- Product code: L2002A3-US
- Lamp Type: Low pressure mercury discharge lamp
- Lamp Dimensions (L x D): 150 mm x 15 mm (5.91" x 0.59")
- Number of Lamps: 4
- Discharge Peaks: 185 nm and 254 nm
- Lamp Lifetime: T80 lifetime 2,000 hours
- Sample UV Intensity: 15±1 mW·cm-2 at 254 nm
- Cleaning Area: 153 mm (6") diameter
- Exposure Time: 1 second – 60 minute, digitally controlled
- Dimensions (W x D x H): 210 mm x 310 mm x 228 mm (8.27" x 12.20" x 8.98")
- Maximum Sample Height: 12 mm (0.47")
- Weight: 5.25 kg
- Case Material: 1.5 mm (0.06") mild steel, textured black powder coating
- Drawer Tray Material: 1.5 mm (0.06") stainless steel, #4 brushed finish
- Display: 4.3" 24 bit color TFT LCD, 480 x 272 px
- Voltage Range: 110 V – 240 V
The cleaner has no integrated ozone filtration system and must be operated in a working fume hood. Because the system runs at high voltage, UV bulb replacement should be carried out by the manufacturer rather than by the user, as incorrect reassembly can be dangerous.
—