The Ossila Electrical Characterization System (T2009F1) is a complete bench-top platform for I-V testing of small-scale electronic devices. It pairs a manual probe station with a dual channel source measure unit, so samples such as photovoltaics, LEDs, OLEDs and transistors can be probed and measured with one setup.
This version uses fine-pitch micrometer micromanipulators with a 5 µm readout and smallest achievable motion of approx. 2-4 µm. The station sits on an aluminum optical breadboard with elastomeric vibration dampening mounts, and its ferromagnetic stainless steel platens hold up to 8 Ossila Micromanipulators. A vacuum chuck with selectable zones accepts substrates from 5 mm to 150 mm diameter, held by an oil-free diaphragm pump with an integrated gauge.
Samples are viewed through a monocular zoom lens and Sony IMX415 4K camera, streaming at up to 60 fps to a 15.6" 1080p monitor on a positionable arm. The SMU combines two source measure channels, two precision voltage sense channels and a shutter/trigger channel, covering -10 V to +10 V across five current ranges.
- Measuring I-V characteristics of photovoltaic devices.
- Electrical testing of LEDs and OLEDs.
- Characterizing transistors and other DC semiconductor devices.
- Triggering or synchronizing with other instruments through the shutter/trigger channel.
| Brand | Ossila |
|---|---|
| CAS number | — |
| Molecular formula | — |
| Purity | — |
| Category | — |
| Pack sizes | — |
| Physical form | — |
| Storage | — |
- Product code: T2009F1
- Footprint (L x W x H): 50 x 35 x 60 cm
- Base: Aluminum optical breadboard, 25 mm hole spacing
- Platen Capacity: Up to 8 Ossila Micromanipulators
- Stage Travel: 25 mm (X, Y), 10 mm (Z), 360° (coarse rotation), 15° (fine rotation)
- Substrate Size: 5 mm to 150 mm diameter
- Chuck Type: Vacuum (selectable zones)
- Chuck Leakage Current: <75 pA at ± 75 VDC
- Vacuum Pump: Diaphragm (oil-free), 7 L/min, -680 mbar (-510 mmHg)
- Microscope: Monocular Zoom Lens (0.6x – 5.0x optical zoom)
- Camera Sensor: Sony IMX415 (4K, 1/2.8")
- Total Magnification: 37x – 309x
- Micrometer Readout Resolution: 5 µm
- Smallest Achievable Motion: Approx. 2-4 µm
- Probe Type: Tungsten, 20 μm tip diameter
- SMU Voltage Range: ±10 V
The vacuum pump is suitable for air and non-flammable, non-corrosive dry gases only (e.g. nitrogen, argon) and exhausts externally; it should not be used with liquids or gases that need to be recovered. Keep the shutter/trigger BNC at 5 V, as higher voltages may damage the port.
—